HOW
TO: run a linearity ramp
-
stop current Run using Calorimeter
Crates
-
ask to stop run and take Calorimeter Crates out
of combined cofiguration
-
set T&C mode
-
open T&C
GUI: /online/config/cal/tanc/TCmon.py
-
set timing for the baseline sampling:
-
click on Global
BLS Mode
-
select x1
- 2 ticks
=> baseline is sampled at -5BC instead
of the nominal -3BC
=> gain path is fixed to x1: this should be used if the ramp reaches high
pulses (DAC > 10000) else use x8 - 2 ticks
-
set ADC Mode
-
click on Global
ADC Mode
-
select PedSub
=> reads out
all cells with Pedestal subtraction but no 0-suppression
-
set T&C Mode
-
click on Global
T&C set
-
select x889
-
set read-out timing w.r.t. puls
start
-
click on T&C
for
crate NW 02 0x42
-
enter 0
in the TICK
field
-
verify T&C setting
-
open Calorimeter
Power Supply Montiroting GUI: /online/config/cal/supplies_all.py
-
go toMode
page
-
T&C Mode
should
be 0x889
-
Diag
should
not be 0 for all crates
-
ADC/BLS Mode
should be 0x8a01 if x1 - 2 ticks or 0x8b01 if x8 - 2ticks
-
set initial Pulser configuration
-
open Pulser
Control GUI: /online/config/cal/pulser.py
-
shutdown all pulsers
-
select all pulsers and click
Shutdown
-
initialize
pulser setting in the GUI
-
select all pulsers you would
like to use
typically all pulsers except ICD (number 12)
-
set initial DAC
Value normaly to
0
-
set channel
enables
set: Enable All
! click twice on the enable all buttons after
a shutdown, all enables must be red
set: Cmd All
! click twice on the enable all buttons after
a shutdown, all enables must be red
for the time being a good value is: Delay
All = 10
! type return, such that the value in the Delay All field is seen in the
6 individual delay fields
-
Download!
all values to pulsers
-
check Pulser settings
-
open Calorimeter
Power Supply Montiroting GUI: /online/config/cal/supplies_all.py
-
go to
PLS
Mode
page:
-
selected channels for all selected
pulsers should be set to 1
-
selected commands for all selected
pulsers should be set to 1
-
DAC values for selected pulsers should
be set to choosen value
-
DELAY values for selected pulsers should
be set to choosen initial value
-
all not selected pulsers should have
DAC value=0, DELAY value=0
-
all not selected channels or commands
should =0
-
start Delay-ramp in the PIB
! in request-field
65 should appear
-
enter ramp parameters
put: 20 (max132000)
put: 5000 for gain x1
put: 1000 for gain x8 (min 1, max 132000)
put: 50
=> ramp between 0 and 100000 (20000) with 20 Steps of DAC 5000 (2000)
units and 50 events per step
-
Download!
ramp
request to PIB
=> a ramp-status window should pop up
-
start a Pulser run
-
open a Taker-Window, or ask DAQ
shifter
select cal-pulsar-test-1.1.xtm
-
start new run
=>
events should start to be counted in the ramp-status window
=>
when a new step of the ramp occurs the delay values in the supply-all GUI
change
-
stop run, when all events are collected
-
write the run-number carefully to
the logbook! and send it to to bassler@in2p3.fr