RUN: 160688 QUALITY: GOOD CRITERIA VERSION: 5 TOTAL NUMBER OF EVENTS: 215 NUMBER OF "ZERO_BIAS" EVENTS: 215 NUMBER OF "UNSUPPRESSED" EVENTS: 0 NUMBER OF "JETS" EVENTS: 0 NUMBER OF "ALL OTHER" EVENTS: 0 --------------------------------------------------------------- ALL towers have a signal. --------------------------------------------------------------- 6.05% of noisy events in total: 0.00% of "ring of fire", 6.05% of "coherent noise", 0.00% of "missing crates" ZERO_BIAS: 6.05% of noisy events: 0.00% of "ring of fire", 6.05% of "coherent noise", 0.00% of "missing crates" --------------------------------------------------------------- 8 "bad" cells These cells are: (ieta,iphi)=(27,21) Layer EM3_a (3) CRATE/ADC/BLS/TOWER/DEPTH = 4/8/3/1/2 E/event: 2.48 RMS: 8.56 occupancy: 0.07 (ieta,iphi)=(32,5) Layer EM3_a (3) CRATE/ADC/BLS/TOWER/DEPTH = 7/8/5/3/2 E/event: 2.36 RMS: 9.29 occupancy: 0.09 HOT! (ieta,iphi)=(-25,54) Layer EM3_b (4) CRATE/ADC/BLS/TOWER/DEPTH = 11/3/6/1/5 E/event: -1.20 RMS: 0.11 occupancy: 0.99 (ieta,iphi)=(26,22) Layer FH1 (11) CRATE/ADC/BLS/TOWER/DEPTH = 4/8/2/2/7 E/event: 1.89 RMS: 5.76 occupancy: 0.05 (ieta,iphi)=(27,21) Layer FH1 (11) CRATE/ADC/BLS/TOWER/DEPTH = 4/8/3/1/7 E/event: 1.64 RMS: 5.49 occupancy: 0.06 (ieta,iphi)=(27,22) Layer FH1 (11) CRATE/ADC/BLS/TOWER/DEPTH = 4/8/3/0/7 E/event: 1.90 RMS: 6.33 occupancy: 0.06 (ieta,iphi)=(33,5) Layer FH4 (14) CRATE/ADC/BLS/TOWER/DEPTH = 7/8/6/0/10 E/event: 2.71 RMS: 8.86 occupancy: 0.06 HOT! (ieta,iphi)=(18,45) Layer CH1 (15) CRATE/ADC/BLS/TOWER/DEPTH = 5/1/6/3/11 E/event: 4.80 RMS: 5.55 occupancy: 0.36 --------------------------------------------------------------- --------------------------------------------------------------- ---> The run quality creteria are: ---> VERSION: 5 ---> If percentage of noisy events > 10.00 the run quality is BAD. ---> else if number of cells with mean energy out of limits > 24 the run quality is BAD. ---> else if number of ZERO_BIAS events in the run < 200 the run quality is UNKNOWN. ---> else if number of towers without signal > 8 the run quality is BAD. ---> else if number of towers without signal > 0 the run quality is REASONABLE. ---> else if number of BAD towers > 0 the run quality is REASONABLE. ---> else if percentage of noisy events > 8.00 the run quality is REASONABLE. ---> else the run quality is GOOD. --------------------------------------------------------------- PARAMETERS IN USE ARE: Mode: OFFLINE Minimal number of events to generate report file: 200 Minimal number of entries per cell: 10 Minimal occupancy per cell to declare it hot: 0.25 Cell mean energy: -1.00 - 1.50 CH cell mean energy: -2.00 - 2.50 Cell RMS: 0.03 - 4.00 Max average occupancy per cell in tower : 0.30 : if abs(ieta) > 32 : 0.50 Max number of bad channels in tower : 3 Tower max mean energy: -0.75 - 0.75